Ram Cannula
Neotech’s RAM Cannula features soft, gently curved prongs designed for patient comfort. Its simple set up and flexible tubing helps promote developmentally appropriate positioning for infants and children. This Neotech RAM Cannula in India is available in seven sizes. All sizes are color-coded and include a 15-millimeter oxygen tubing adapter for use with low-flow or high flow nasal cannula.
Step 1 Select proper size nasal oxygen cannula by using RAM size guide provided in package. Ideal prongs size will fill approximately 80% of nares. Ensure that prongs do not fill nares completely.
Step 2 Attach 15 mm oxygen tubing adapter (provided) to Neotech RAM cannula. Attach cannula to oxygen source. Ensure all cannula connections are properly secured.
Step 3 Open and test the level of oxygen flow before putting the device on the patient.
Step 4 Insert prongs into patient’s nares and allow a small gap between patient’s septum and base of prongs.
Step 5 Secure cannula to face with EZ-Hold® Tubing and Cannula Holder or comparable product.
Step 6 Replace Nasal Oxygen Cannula every 14 days or per hospital protocol, whichever is sooner. Dispose of used Neotech RAM Cannula per hospital protocol.
INDICATIONS FOR USE
The Neotech RAM Cannula in India is intended to be used for preterm and term neonates, infants, and pediatric patients who require supplemental oxygen, respiratory support or assistance in breathing in an ambulatory, hospital or institutional environment. It is used to provide oxygen with the following methods:
1. Heated or unheated, humidified low flow oxygen.
2. Heated, humidified high flow oxygen.
CONTRAINDICATIONS, CAUTIONS & WARNINGS
The Neotech RAM Cannula in India is contraindicated for patients with nasal atresia or patients with facial structure deformities that prohibit adequate respiratory support.
The Nasal Cannula is contraindicated for patients with nasal atresia or patients with facial structure deformities that prohibit adequate respiratory support.
Federal law restricts this device to sale by or on the order of a physician.
This device should only be used while the patient is under the continuous, direct supervision of healthcare professional and frequent observation of prongs position in patient’s nares is necessary
Improper selection of size, improper positioning or improper use may result in septal trauma or necrosis.
Cannula tubing can pose a potential strangulation hazard. Do not use the O-ring to secure the device to the patient.
Always begin gas flow prior to inserting prongs into patient’s nares.
Do not use if package is damaged or open.
Do not soak, wash or sterilize.
Discontinue immediately if skin irritation occurs.
Single patient use only.
Cat No. | Color/Item | Qty/Unit |
N4900 | White | 10/box |
N4901 | Green | 10/box |
N4902 | Blue | 10/box |
N4903 | Orange | 10/box |
With Larger Tubing | ||
N4904 | Yellow | 10/box |
N4905 | Teal | 10/box |
N4906 | Purple | 10/box |
N4820 | Oxygen Adapter | 50/box |